Name:

Patient monitor SPR-9000

 

Name:

Patient monitor SPR-9000

Code:MST118303
PRODUCTS INTRODUCTION:

Features &Benefits:
6standard parameters: ECG,RESP,NIBP,SPO2,1-TEMP,PR/HR.
12.1”high resolution color TFT LTD display..
Real time S-T segment analysis,pace-maker detection and ARR analysis.
Drug calculation and titration table.
Multidisplay selectable  including standard,large font,trend coexis,OxyCRG dynamic view and bed-to-bed view display.
Indication of Incorrect manipulation and malfunction analysis.
On-line help and Patient info input management
Multi-lead ECG waveforms display in phase.
Large volume of tabular and graphic trends information storage and easy to recall
Capture dynamic waveforms.
Efficient resistance to interference of defibrillator and HF knife.
Up to 4 hours working capacity of built-in rechargeable battery.
Networking capacity and TCP/IP networking platform.
Option of IBP and EtCO2

Specifications
 
Performance Specifications
Display:12.1’’color TFT
Rolling and refreshing waveform display
Resolution:800×600
Multi display selectable including:
Large-font display
Trend coexist display
OxyCRG dynamic view display.
Bed-to-bed view display
Trace:9 waveforms(7 ECG, 1 SPO2 and 1 RESP)
Sweep speed:12.5mm/s,25mm/s,50mm/s
Indicator:Power/battery indicator light
QRS beep and alarm sound
Battery:Rechargeable lead acid cell,12v/4AH
Max.24 hour for charging,4 hours for continue working
Trend time:Above 72 hours
Alarm:User-adjustable High,Medium and Low limits 3-level Audible and visual alarm
Networking:Connected to central monitoring system
Tcp/Ip netting platform
Recorder:Build-in ,thermal array
Plethysmogram caveform:2 channels
Record mode: manual, on alarm, time-defined
Recording width:50mm
Printing speed:50mm/s
Recording type:Frozen waveform record
NIBP recall record
Trend table record
Alarm record
Fixed -time record

ECG
Lead mode:5 –lead(R,L,F,N,C)
Lead selection :I,II,III,avR,avL,avF,V
Waveform:     2 and 7 channel selectable
Gain selection: 0.25mm/mv,0.5mm/mv,1mm/mv,2mm/mv
Sweep speed: 12.5mm/s;25mm/s;50mm/s
Heart rate range: Adult:15~300bpm;Neonate:/pediatric:15~350bpm
Accuracy: +1bpm or +1%,whichecer is greater
Resolution:     1bpm
Filter:surgery mode:1~20Hz
monitor model:0.5~40Hz
Diagnostic mode:0.05~130Hz
Scaling signal: 1mv,+3%
Protection:      withstand 4000VAC/50 voltage isolation against electrosurgical interference and defibrillation
Alarm range: 15~350bpm
S-T segment detecton:
Measurement range:  2.0mV~+2.0mV                        
 
Alarm Range:-2.0mV~ +2.0mV
Accuracy:-0.8mV ~+0.8Mv
Error:+0.02Mv
 
Arrhythmia analysis:YES

NIBP
Method:Digital Automatic oscillometric
Operation mode :Manual/Automatic/continuous
Auto measurement time:Adjustable(1~480min)
Measurement Unit:mmHg/Kpa selectable
Measurement types:Systolic,Diastolic,Mean 
Measurement tange:
Range of Systolic pressure:Adult:40~270mmHg
Pediatric:40~220mmHg
Neonate:40~135mmHg
Range of mean pressure:Adult:20~235mmHg
Pediatric:20~165mmHg
Neonate:20~110mmHg
Range of diastolic pressure:Adult:10~215mmHg
Pediatric:10~150mmHg
Neonate:10~100mmHg
Over-pressure protection:Double safety protection
Resolution:1 mmHg
Alarm:Systolic.Diastolic,Mean

SPO2
Measurement range:0~100%
Resolution:1%
Accuracy:+2%(70-100%);0-69% unspecified
Alarm range0~100%
Pluse Rate:range:20~300bpm
Resolution:1bpm
Error:+1bpm or +2%,whichever is greater

RESPIRATION
Method:Thoracic impedance
Measurement range:
Adult:7~120rpm;
Neonate/Pediatric:7~150rpm
Apnea alarm:YES,10~40s
Resolution:1rpm
Accuracy:+2rpm

TEMPERATURE
Compatible probe:YSI or CYF
Measurement range:5~50℃
Resolution :0.1℃
Accuracy:+0.1℃
Refreshing time:about 1
Average measuring time:<10s

EtCO2
Sidertream CO2
Measurement range:0~99mmHg
Accuracy:+2mmHg(0~40mmHg)
Sampling Range:100ml/min
Sampling rate accuracy:15%
Respiration rate0~120rmp
Respiration accuracy:+2rmp(0~70rmp)
+5rmp(>70rmp)
Respiration time:<240msec(10% to 90%)
Delay time:<2s

Standard Configuration:ECG, NIBP,RESP,1-TEMP,SPO2,PR
Optional:Thermal Recorder,1-IBP,2-IBP,ETCO2 

Miscellaneous
 
Satety:             
Safety level: Class I, type CF

Dimendion and Weight
Dimendion:440×430×450         
G.Weight:<9.0KS

Operation Environment
Temperature:Working 0~+40℃
Transportation and storage –20~+60℃
Humidity:working≤85%
Transportation and storage≤93%
Power:AC 100-240,50/60Hz
Patient range:Neonate,pediatric,and adult patients

Accessories supplies
(1) 5 lead ECG cable
(2) 1 spo2 probes
(3) 1 NIBP prbe
(4) 1 temp probe
(5)1 ground connecting liner
(6) Chest electrode (10pcs/set)

Contact Information:
 Guangdong Shantou Minston Medical Instruments Co., Ltd.
 Tel:86-754-88481863
 E-mail:wuyubin1011@gmail.com
 http://www.medical-products-cn.com
COPY RIGHT(C)2007